Ultrasound in Medical Diagnostics

Ultrasound in Medical Diagnostics

When used for imaging, ultrasonic waves are emitted from a transducer, a crystal exhibiting the piezoelectric effect (the expansion and contraction of a substance when a voltage is applied across it, causing a vibration of the crystal). These high-frequency vibrations are transmitted into any tissue in contact with the transducer. Similarly, if a pressure is applied to the crystal (in the form of a wave reflected off tissue layers), a voltage is produced which can be recorded. The crystal therefore acts as both a transmitter and a receiver of sound. Ultrasound is also partially absorbed by tissue on its path, both on its journey away from the transducer and on its return journey. From the time between when the original signal is sent and when the reflections from various boundaries between media are received, (as well as a measure of the intensity loss of the signal), the nature and position of each boundary between tissues and organs may be deduced.

Reflections at boundaries between two different media occur because of differences in a characteristic known as the acoustic impedance \(Z\) of each substance. Impedance is defined as

\(Z=\text{ρv},\)

where \(\rho \) is the density of the medium (in \({kg/m}^{3}\)) and \(v\) is the speed of sound through the medium (in m/s). The units for \(Z\) are therefore \({kg/(m}^{2}·\text{s)}\).

This table shows the density and speed of sound through various media (including various soft tissues) and the associated acoustic impedances. Note that the acoustic impedances for soft tissue do not vary much but that there is a big difference between the acoustic impedance of soft tissue and air and also between soft tissue and bone.

The Ultrasound Properties of Various Media, Including Soft Tissue Found in the Body

At the boundary between media of different acoustic impedances, some of the wave energy is reflected and some is transmitted. The greater the difference in acoustic impedance between the two media, the greater the reflection and the smaller the transmission.

The intensity reflection coefficient \(a\) is defined as the ratio of the intensity of the reflected wave relative to the incident (transmitted) wave. This statement can be written mathematically as

\(a=\cfrac{{({Z}_{2}-{Z}_{1})}^{2}}{{({Z}_{1}+{Z}_{2})}^{2}},\)

where \({Z}_{1}\) and \({Z}_{2}\) are the acoustic impedances of the two media making up the boundary. A reflection coefficient of zero (corresponding to total transmission and no reflection) occurs when the acoustic impedances of the two media are the same. An impedance “match” (no reflection) provides an efficient coupling of sound energy from one medium to another. The image formed in an ultrasound is made by tracking reflections (as shown in this figure) and mapping the intensity of the reflected sound waves in a two-dimensional plane.

Example: Calculate Acoustic Impedance and Intensity Reflection Coefficient: Ultrasound and Fat Tissue

(a) Using the values for density and the speed of ultrasound given in this table, show that the acoustic impedance of fat tissue is indeed \(1.34×{10}^{6}\phantom{\rule{0.25em}{0ex}}{kg/(m}^{2}·s)\).

(b) Calculate the intensity reflection coefficient of ultrasound when going from fat to muscle tissue.

Strategy for (a)

The acoustic impedance can be calculated using \(Z=\text{ρv}\) and the values for \(\rho \) and \(v\) found in this table.

Solution for (a)

(1) Substitute known values from this table into \(Z=\text{ρv}\).

\(Z=\mathrm{\rho v}=(\text{925 kg}{\text{/m}}^{3})(\text{1450 m/s})\)

(2) Calculate to find the acoustic impedance of fat tissue.

\(1.34×{10}^{6}\phantom{\rule{0.25em}{0ex}}{kg/(m}^{2}·s)\)

This value is the same as the value given for the acoustic impedance of fat tissue.

Strategy for (b)

The intensity reflection coefficient for any boundary between two media is given by \(a=\cfrac{{({Z}_{2}-{Z}_{1})}^{2}}{{({Z}_{1}\phantom{\rule{0.25em}{0ex}}+\phantom{\rule{0.25em}{0ex}}{Z}_{2})}^{2}}\), and the acoustic impedance of muscle is given in this table.

Solution for (b)

Substitute known values into \(a=\cfrac{{({Z}_{2}-{Z}_{1})}^{2}}{{({Z}_{1}\phantom{\rule{0.25em}{0ex}}+\phantom{\rule{0.25em}{0ex}}{Z}_{2})}^{2}}\) to find the intensity reflection coefficient:

\(a=\cfrac{{({Z}_{2}-{Z}_{1})}^{2}}{{({Z}_{1}+{Z}_{2})}^{2}}=\cfrac{{(1\text{.}\text{34}×{\text{10}}^{6}\phantom{\rule{0.25em}{0ex}}{\text{kg/(m}}^{\text{2}}\mathrm{· s)}-1.70×{\text{10}}^{6}\phantom{\rule{0.25em}{0ex}}{\text{kg/(m}}^{\text{2}}· s))}^{2}}{{(1\text{.}\text{70}×{\text{10}}^{6}\phantom{\rule{0.25em}{0ex}}{\text{kg/(m}}^{\text{2}}· s)+1\text{.}\text{34}×{\text{10}}^{6}\phantom{\rule{0.25em}{0ex}}{\text{kg/(m}}^{\text{2}}· s))}^{2}}=0\text{.}\text{014}\)

Discussion

This result means that only 1.4% of the incident intensity is reflected, with the remaining being transmitted.

The applications of ultrasound in medical diagnostics have produced untold benefits with no known risks. Diagnostic intensities are too low (about \({\text{10}}^{-2}\phantom{\rule{0.25em}{0ex}}{\text{W/m}}^{2}\)) to cause thermal damage. More significantly, ultrasound has been in use for several decades and detailed follow-up studies do not show evidence of ill effects, quite unlike the case for x-rays.

The most common ultrasound applications produce an image like that shown in this figure. The speaker-microphone broadcasts a directional beam, sweeping the beam across the area of interest. This is accomplished by having multiple ultrasound sources in the probe’s head, which are phased to interfere constructively in a given, adjustable direction. Echoes are measured as a function of position as well as depth. A computer constructs an image that reveals the shape and density of internal structures.

How much detail can ultrasound reveal? The image in this figure is typical of low-cost systems, but that in this figure shows the remarkable detail possible with more advanced systems, including 3D imaging. Ultrasound today is commonly used in prenatal care. Such imaging can be used to see if the fetus is developing at a normal rate, and help in the determination of serious problems early in the pregnancy. Ultrasound is also in wide use to image the chambers of the heart and the flow of blood within the beating heart, using the Doppler effect (echocardiology).

Whenever a wave is used as a probe, it is very difficult to detect details smaller than its wavelength \(\lambda \). Indeed, current technology cannot do quite this well. Abdominal scans may use a 7-MHz frequency, and the speed of sound in tissue is about 1540 m/s—so the wavelength limit to detail would be \(\lambda =\cfrac{{v}_{w}}{f}=\cfrac{\text{1540 m/s}}{7×{\text{10}}^{6}\phantom{\rule{0.25em}{0ex}}\text{Hz}}=0\text{.}\text{22 mm}\). In practice, 1-mm detail is attainable, which is sufficient for many purposes. Higher-frequency ultrasound would allow greater detail, but it does not penetrate as well as lower frequencies do. The accepted rule of thumb is that you can effectively scan to a depth of about \(500\lambda \) into tissue. For 7 MHz, this penetration limit is \(500×0.22\text{ mm}\), which is 0.11 m. Higher frequencies may be employed in smaller organs, such as the eye, but are not practical for looking deep into the body.

In addition to shape information, ultrasonic scans can produce density information superior to that found in X-rays, because the intensity of a reflected sound is related to changes in density. Sound is most strongly reflected at places where density changes are greatest.

Another major use of ultrasound in medical diagnostics is to detect motion and determine velocity through the Doppler shift of an echo, known as Doppler-shifted ultrasound. This technique is used to monitor fetal heartbeat, measure blood velocity, and detect occlusions in blood vessels, for example. (See this figure.) The magnitude of the Doppler shift in an echo is directly proportional to the velocity of whatever reflects the sound. Because an echo is involved, there is actually a double shift. The first occurs because the reflector (say a fetal heart) is a moving observer and receives a Doppler-shifted frequency. The reflector then acts as a moving source, producing a second Doppler shift.

A clever technique is used to measure the Doppler shift in an echo. The frequency of the echoed sound is superimposed on the broadcast frequency, producing beats. The beat frequency is \({F}_{B}=\mid {f}_{1}-{f}_{2}\mid \), and so it is directly proportional to the Doppler shift (\({f}_{1}-{f}_{2}\)) and hence, the reflector’s velocity. The advantage in this technique is that the Doppler shift is small (because the reflector’s velocity is small), so that great accuracy would be needed to measure the shift directly. But measuring the beat frequency is easy, and it is not affected if the broadcast frequency varies somewhat. Furthermore, the beat frequency is in the audible range and can be amplified for audio feedback to the medical observer.

Uses for Doppler-Shifted Radar

Doppler-shifted radar echoes are used to measure wind velocities in storms as well as aircraft and automobile speeds. The principle is the same as for Doppler-shifted ultrasound. There is evidence that bats and dolphins may also sense the velocity of an object (such as prey) reflecting their ultrasound signals by observing its Doppler shift.

Example: Calculate Velocity of Blood: Doppler-Shifted Ultrasound

Ultrasound that has a frequency of 2.50 MHz is sent toward blood in an artery that is moving toward the source at 20.0 cm/s, as illustrated in this figure. Use the speed of sound in human tissue as 1540 m/s. (Assume that the frequency of 2.50 MHz is accurate to seven significant figures.)

  1. What frequency does the blood receive?
  2. What frequency returns to the source?
  3. What beat frequency is produced if the source and returning frequencies are mixed?

Industrial and Other Applications of Ultrasound

Industrial, retail, and research applications of ultrasound are common. A few are discussed here. Ultrasonic cleaners have many uses. Jewelry, machined parts, and other objects that have odd shapes and crevices are immersed in a cleaning fluid that is agitated with ultrasound typically about 40 kHz in frequency. The intensity is great enough to cause cavitation, which is responsible for most of the cleansing action. Because cavitation-produced shock pressures are large and well transmitted in a fluid, they reach into small crevices where even a low-surface-tension cleaning fluid might not penetrate.

Sonar is a familiar application of ultrasound. Sonar typically employs ultrasonic frequencies in the range from 30.0 to 100 kHz. Bats, dolphins, submarines, and even some birds use ultrasonic sonar. Echoes are analyzed to give distance and size information both for guidance and finding prey. In most sonar applications, the sound reflects quite well because the objects of interest have significantly different density than the medium in which they travel. When the Doppler shift is observed, velocity information can also be obtained. Submarine sonar can be used to obtain such information, and there is evidence that some bats also sense velocity from their echoes.

Similarly, there are a range of relatively inexpensive devices that measure distance by timing ultrasonic echoes. Many cameras, for example, use such information to focus automatically. Some doors open when their ultrasonic ranging devices detect a nearby object, and certain home security lights turn on when their ultrasonic rangers observe motion. Ultrasonic “measuring tapes” also exist to measure such things as room dimensions. Sinks in public restrooms are sometimes automated with ultrasound devices to turn faucets on and off when people wash their hands. These devices reduce the spread of germs and can conserve water.

Ultrasound is used for nondestructive testing in industry and by the military. Because ultrasound reflects well from any large change in density, it can reveal cracks and voids in solids, such as aircraft wings, that are too small to be seen with x-rays. For similar reasons, ultrasound is also good for measuring the thickness of coatings, particularly where there are several layers involved.

Basic research in solid state physics employs ultrasound. Its attenuation is related to a number of physical characteristics, making it a useful probe. Among these characteristics are structural changes such as those found in liquid crystals, the transition of a material to a superconducting phase, as well as density and other properties.

These examples of the uses of ultrasound are meant to whet the appetites of the curious, as well as to illustrate the underlying physics of ultrasound. There are many more applications, as you can easily discover for yourself.

Check Your Understanding

Why is it possible to use ultrasound both to observe a fetus in the womb and also to destroy cancerous tumors in the body?

Solution

Ultrasound can be used medically at different intensities. Lower intensities do not cause damage and are used for medical imaging. Higher intensities can pulverize and destroy targeted substances in the body, such as tumors.